Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The choice of an appropriate objective function for calibration depends on the intended application of the model. Different objective functions are designed with the intention of emphasizing the fit of modelled flow to different aspects of the observed hydrograph (Vaze et al., 2011). The objective functions available in the Source Calibration Wizard are listed in Table 1, including useful references for further information. Refer to the Bivariate Statistics User Guide entry for the objective function equations and further information on their interpretation. 

Anchor
Table 1
Table 1
Table 1. List of Source calibration objective functions.

Objective Function NameDescriptionReference
NSE DailyMaximise the NSE of daily flowsVaze et al. (2011), Section 6
NSE MonthlyMaximise the NSE of monthly flowsVaze et al. (2011), Section 6
NSE Log DailyMaximise the NSE of the logarithm of daily flows
Minimise Absolute BiasMinimise the Absolute value of the relative biasVaze et al. (2011), Section 6

NSE Daily & Bias PenaltyMaximise the NSE of daily flows and bias penaltyVaze et al. (2011), Section 6
NSE Log Daily & Bias PenaltyMaximise the NSE of the logarithm of daily flows and bias penalty
NSE Monthly & Bias PenaltyMaximise the NSE of monthly flows and bias penaltyVaze et al. (2011), Section 6
NSE Daily & Flow DurationMaximise the NSE of daily flows and the NSE of the flow durationVaze et al. (2011), Section 6
NSE Daily & Log Flow DurationMaximise the NSE of daily flows and the NSE of the flow duration of log flowsVaze et al. (2011), Section 6
Square-root Daily, Exceedance and Bias

Minimise a combination of the bias, daily Flows and daily exceedance (flow duration) curve

Lerat et al., 2013

Implementation Details

The Bivariate Statistics SRG User Guide entry provides general information on the objective function equations and their interpretation. Implementation details that are specific to the Source Calibration Wizard are described below.

Missing Data

It is common for observed time series of hydrological processes to contain missing values. Also, the observed and modelled time series may have different start and end dates. The Source calibration tool calculates the objective function values using only data from those time steps for which both observed and modelled data is available. In other words, the calibration objective function are calculated using observed and modelled data has been filtered to include only:

  • data from within the calibration period, and
  • data for time steps with complete data pairs.

Monthly Flows

The NSE Monthly objective function uses monthly streamflow values. These are calculated as follows:

  • If the model is run on a daily time step, monthly flows are calculated by summing the observed and modelled daily flow values. The NSE calculation ignores observed and modelled data for all months where there are one or more days of missing data in the observed flow series. 
  • If the model is run on a monthly time step, then the monthly values are unchanged.

Data

Input data

Details on data to be input by the modeller are provided in the Source User Guide. Requirements for data series inputs to the various objective functions are included in the descriptions of each objective function, above.

...

Modellers have the option of selecting one optimisation technique, two optimisation techniques (in series), or manual optimisation.   Modellers can also select which objective function they wish to use.   The other parameters the modeller can input are described in the following table:in Table 2

Anchor
Table 2
Table 2
Table 2. Objective function parameters. Refer to the Bivariate Statistics SRG User Guide entry for further information on their implementation.

Objective Function
ParameterParameter DescriptionUnitsDefaultRange
NSE Daily & Flow DurationaWeight on NSE in the combined objectiveDimensionless0.50 ≤ α ≤ 1
NSE Daily & Log Flow DurationaWeight on NSE in the combined objectiveDimensionless0.50 ≤ α ≤ 1

...